Name the "filename" at the top, make sure it is named XXXXXX0001 so when you save, it will automatically increase the last digit to XXXXX0002
In the "Drive and Folder", select "Ex6". Then in the space below, type in the new folder name. Then click "create folder". In the "Drive and Folder", click the folder you just created
Then click "Select New Folder" and "Ok"
At very top of the screen you should see the directory with the file name that ends with XXXXX0001 that you just created. If not, go back and check
4.Operate> Scan Spread
This will make sure there are 4 boxes (2 data logger and 2 batteries). Then click "OK" to apply the setting
If this show just only 2 or 0, go back and check the connection.
5.Test> Geophone resistance Click "test"
To make sure there are 24 operational geophones, you should see 24 bars of the same height. If some bars are not of that height, check the connection of that geophone.
Look at the lower left corner, it will arm and then start recording the signal
7. File> Save
Make sure to do this step, so the file will save and automatically the last digit of the filename will increase by 1
8. Repeat Step 6 till you have more than 8 records
For Seismic Refraction Experiments:
1. Open the program "Ex-6"
2. Set parameters
File> Load Confinguration
Select file "Campusfieldexperiment". Click "Open"
To double check the parameters:
Setup> Acquistion
Sample rate: 1ms
Stack count: 2
Preamp Gain: 24 db
Auto Arm: Off
Auto Save: On
Inc File After Save: On
Record Length: 1.00 sec
Trigger at any box: On
Trigger Sensitivity: Low
Hammer Switch: On
Pretrigger: On : 26.00ms
Compress Data: On
3. Set file directory
File>File name
Name the "filename" at the top, make sure it is named XXXXXX0001 so when you save, it will automatically increase the last digit to XXXXX0002
In the "Drive and Folder", select "Ex6". Then in the space below, type in the new folder name. Then click "create folder". In the "Drive and Folder", click the folder you just created
Then click "Select New Folder" and "Ok"
At very top of the screen you should see the directory with the file name that ends with XXXXX0001 that you just created. If not, go back and check
4.Operate> Scan Spread
This will make sure there are 4 boxes (2 data logger and 2 batteries). Then click "OK" to apply the setting
If this show just only 2 or 0, go back and check the connection.
5.Test> Geophone resistance Click "test"
To make sure there are 24 operational geophones, you should see 24 bars of the same height. If some bars are not of that height, check the connection of that geophone.
For ReMi experiments:
1. Open the program "Ex-6"
2. Set parameters
File> Load Confinguration
Select file "Ex6ReMIV1". Click "Open"
To double check the parameters: Setup> Acquistion
The configuration parameters for a ReMi study should look like this:
Sample rate: 2ms
Stack count: 1
Preamp Gain: 24 db
Auto Arm: Off
Auto Save: On
Inc File After Save: On
Record Length: 32.00 sec
Trigger at any box: On
Trigger Sensitivity: Low
Hammer Switch: On
Pretrigger: Off
Compress Data: On
3. Set file directory
File>File name
Name the "filename" at the top, make sure it is named XXXXXX0001 so when you save, it will automatically increase the last digit to XXXXX0002
In the "Drive and Folder", select "Ex6". Then in the space below, type in the new folder name. Then click "create folder". In the "Drive and Folder", click the folder you just created
Then click "Select New Folder" and "Ok"
At very top of the screen you should see the directory with the file name that ends with XXXXX0001 that you just created. If not, go back and check
4.Operate> Scan Spread
This will make sure there are 4 boxes (2 data logger and 2 batteries). Then click "OK" to apply the setting
If this show just only 2 or 0, go back and check the connection.
5.Test> Geophone resistance Click "test"
To make sure there are 24 operational geophones, you should see 24 bars of the same height. If some bars are not of that height, check the connection of that geophone.
*The program may freeze when you click test. Just close the program and start again. If you try more than 3 times, make a note and skip this step.
6. Operate> Trigger System
Look at the lower left corner, it will arm and then start recording the signal
7. File> Save
Make sure to do this step, so the file will save and automatically the last digit of the filename will increase by 1
8. Repeat Step 6 till you have more than 8 records
For Seismic Refraction Experiments:
1. Open the program "Ex-6"
2. Set parameters
File> Load Confinguration
Select file "Campusfieldexperiment". Click "Open"
To double check the parameters:
Setup> Acquistion
Sample rate: 1ms
Stack count: 2
Preamp Gain: 24 db
Auto Arm: Off
Auto Save: On
Inc File After Save: On
Record Length: 1.00 sec
Trigger at any box: On
Trigger Sensitivity: Low
Hammer Switch: On
Pretrigger: On : 26.00ms
Compress Data: On
3. Set file directory
File>File name
Name the "filename" at the top, make sure it is named XXXXXX0001 so when you save, it will automatically increase the last digit to XXXXX0002
In the "Drive and Folder", select "Ex6". Then in the space below, type in the new folder name. Then click "create folder". In the "Drive and Folder", click the folder you just created
Then click "Select New Folder" and "Ok"
At very top of the screen you should see the directory with the file name that ends with XXXXX0001 that you just created. If not, go back and check
4.Operate> Scan Spread
This will make sure there are 4 boxes (2 data logger and 2 batteries). Then click "OK" to apply the setting
If this show just only 2 or 0, go back and check the connection.
5.Test> Geophone resistance Click "test"
To make sure there are 24 operational geophones, you should see 24 bars of the same height. If some bars are not of that height, check the connection of that geophone.
*The program may freeze when you click test. Just close the program and start again. If you try more than 3 times, make a note and skip this step.
6. Operate> Arm
Look at the lower left corner "Armed Waiting for Trigger" will appear.
7. Use Hammer or other active source with correctly attached and connected trigger cable to begin the survey.
8. File> Save
Make sure to do this step, so the file will save and automatically the last digit of the filename will increase by 1
name will increase by 1